Adverse Reactions

ADVERSE REACTIONS Several adverse reactions due to technetium Tc 99m medronate have been reported. These were usually hypersensitivity reactions characterized by itching, various skin rashes, hypotension, chills, nausea and vomiting. There have also been rare cases of dizziness and asthenia associated with the use of technetium Tc 99m medronate.

Contraindications

CONTRAINDICATIONS None known.

Description

DESCRIPTION Each reaction vial contains a sterile, nonpyrogenic, nonradioactive lyophilized mixture of 20 mg medronic acid, 1 mg ascorbic acid, 0.13 mg (minimum) stannous fluoride, SnF 2 and 0.38 mg total tin, maximum (as stannous fluoride, SnF 2 ). The pH is adjusted with sodium hydroxide or hydrochloric acid to 6.5 (6.3 to 6.7) prior to lyophilization. The vial does not contain a preservative. The contents of the vial are lyophilized and sealed under nitrogen at the time of manufacture. The pH of the reconstituted product is 5.4 to 6.8. The structure of medronic acid is given below: The precise structure of technetium Tc 99m medronate is unknown at this time. When sterile, pyrogen-free sodium pertechnetate Tc 99m injection is added to the vial, the diagnostic agent technetium Tc 99m medronate is formed for administration by intravenous injection. Structure of Medronic Acid

Dosage And Administration

DOSAGE AND ADMINISTRATION After preparation with oxidant-free sodium pertechnetate Tc 99m injection the suggested dose range of technetium Tc 99m medronate injection in the average patient (70 kg) is 370 MBq to 740 MBq (10 mCi to 20 mCi) given intravenously. Imaging post injection is optimal at 1 to 4 hours. The patient dose should be measured by a suitable radioactivity calibration system immediately prior to administration. Parenteral drug products should be inspected visually for particulate matter and discoloration prior to administration whenever solution and container permit. Shielding should be utilized when preparing the technetium Tc 99m medronate injection. Radiation Dosimetry The effective half-life was assumed to be the physical half-life for all calculated values. The estimated radiation absorbed doses to an average patient (70 kg) from an intravenous injection of a maximum dose of 740 MBq (20 mCi) of technetium Tc 99m medronate are shown in Table 4 . Table 4 Absorbed Radiation Dose Organ mGy/740 MBq rads/20 mCi Total Body 1.3 0.13 Bone Total 7.0 0.70 Red Marrow 5.6 0.56 Kidneys 8.0 0.80 Liver 0.6 0.06 Bladder Wall 2-hr. void 26.0 2.60 4.8-hr. void 62.0 6.20 Ovaries 2-hr. void 2.4 0.24 4.8-hr. void 3.4 0.34 Testes 2-hr. void 1.6 0.16 4.8-hr. void 2.2 0.22 Method of Calculation: “S” Absorbed Dose per Unit Cumulated Activity for Selected Radionuclides and Organs, MIRD Pamphlet No. 11, 1975

Indications And Usage

INDICATIONS AND USAGE Technetium Tc 99m medronate may be used as a bone imaging agent to delineate areas of altered osteogenesis.

Warnings

WARNINGS This class of compounds is known to complex cations such as calcium. Particular caution should be used with patients who have or who may be predisposed to hypocalcemia (i.e., alkalosis). Preliminary reports indicate interference with brain scans using sodium pertechnetate Tc 99m injection which have been preceded by a bone scan using an agent containing stannous ions. This interference may result in false-positive or false-negative brain scans. It is recommended, where feasible, that brain scans precede bone imaging procedures.

Clinical Pharmacology

CLINICAL PHARMACOLOGY During the initial 24 hours following intravenous injection of technetium Tc 99m medronate, about 50 percent of the dose is retained in the skeleton, and about 50 percent is excreted in the urine. A minimum amount of uptake has been observed in soft-tissue organs, most notably the kidneys. Clearance of radioactivity from the blood is quite rapid, with about 10 percent of the injected dose remaining at one hour, and less than 5 and 2 percent at two and four hours, respectively. The rapid blood clearance (T ½: 38 to 75 minutes) provides bone to non-osseous tissue ratios favoring early imaging. Following intravenous administration of technetium Tc 99m medronate, skeletal uptake occurs as a function of blood flow to bone and bone efficiency in extracting the complex. Bone mineral crystals are generally considered to be hydroxyapatite, and the complex appears to have an affinity for the hydroxyapatite crystals in bone. Deposition of radioactivity in bone is rapid and appears to be related to osteogenic activity as well as the aforementioned skeletal blood perfusion. Skeletal uptake is bilaterally uniform, with larger concentrations in the axial structure and in the long bones. Increased accumulation of radioactivity may be seen, generally, in any bone disease state in which there is increased osteogenesis or a localized increase in osseous blood perfusion; consequently, bone imaging agents generally are not effective in detecting chronic bone diseases.

Effective Time

20210427

Version

2

Dosage And Administration Table

Table 4

Absorbed Radiation Dose

Organ

mGy/740 MBq

rads/20 mCi

Total Body

1.3

0.13

Bone Total

7.0

0.70

Red Marrow

5.6

0.56

Kidneys

8.0

0.80

Liver

0.6

0.06

Bladder Wall

2-hr. void

26.0

2.60

4.8-hr. void

62.0

6.20

Ovaries

2-hr. void

2.4

0.24

4.8-hr. void

3.4

0.34

Testes

2-hr. void

1.6

0.16

4.8-hr. void

2.2

0.22

Spl Product Data Elements

Kit for the Preparation of Technetium Tc 99m Medronate technetium tc 99m medronate TECHNETIUM TC-99M MEDRONATE TECHNETIUM TC-99M MEDRONATE MEDRONIC ACID ASCORBIC ACID STANNOUS FLUORIDE SODIUM HYDROXIDE HYDROCHLORIC ACID

Carcinogenesis And Mutagenesis And Impairment Of Fertility

Carcinogenesis, Mutagenesis, Impairment of Fertility No long-term animal studies have been performed to evaluate carcinogenic potential or whether technetium Tc 99m medronate affects fertility in males or females.

Application Number

NDA018107

Brand Name

Kit for the Preparation of Technetium Tc 99m Medronate

Generic Name

technetium tc 99m medronate

Product Ndc

65857-505

Product Type

HUMAN PRESCRIPTION DRUG

Route

INTRAVENOUS

Package Label Principal Display Panel

PRINCIPAL DISPLAY PANEL - Carton Label Carton Label DIAGNOSTIC NDC 65857-505-10 Kit for the Preparation of Technetium Tc 99m Medronate Rx only • Store the product at 20° to 25°C (68° to 77°F) • For Intravenous Use after radiolabeling with sodium pertechnetate Tc 99m. After reconstitution, store refrigerated 2° to 8°C (36° to 46°F) • Use within 6 hours after reconstitution. Contents: 10 sterile multi-dose reaction vials 20 radioactivity labels Prescribing information Cardinal Health™ NONRADIOACTIVE Kit for the Preparation of Technetium Tc 99m Medronate For preparation of technetium Tc 99m medronate for bone imaging Each reaction vial contains a sterile, nonpyrogenic nonradioactive lyophilized mixture of 20 mg medronic acid , 1 mg ascorbic acid, and 0.13 mg (minimum) stannous fluoride SnF 2 and 0.38 mg total tin, maximum (as stannous fluoride, SnF 2 ); pH adjusted with sodium hydroxide or hydrochloric acid prior to lyophilization. Distributed by Cardinal Health 414, LLC Dublin, OH 43017 Contains no preservative. Vial contents are sealed under nitrogen at time of manufacture. Recommended Dose: See Prescribing Information 876659-H01 LOT 12345 YYY-MM-DD Principal Display Panel - Carton Label

Spl Unclassified Section

PHYSICAL CHARACTERISTICS Technetium Tc 99m decays by isomeric transition with a physical half-life of 6.02 hours. 1 The principal photon that is useful for detection and imaging studies is shown in Table 1 . Table 1 1 Kocher, David C: Radioactive Decay Data Tables, DOE/TlC-11026, 108, 1981 Principal Radiation Emission Data Radiation Mean % per Disintegration Mean Energy (keV) Gamma-2 89.07 140.5 External Radiation The specific gamma ray constant for Tc 99m is 0.78 R/ hour-millicurie at 1 cm. The first half-value layer is 0.017 cm lead (Pb). A range of values for the relative attenuation of the radiation emitted by this radionuclide that results from interposition of various thicknesses of Pb is shown in Table 2 . To facilitate control of the radiation exposure from millicurie amounts of this radionuclide, the use of a 0.25 cm thickness of Pb will attenuate the radiation emitted by a factor of about 1,000. Table 2 Radiation Attenuation by Lead Shielding Shield Thickness (Pb) cm Coefficient of Attenuation 0.017 0.5 0.08 10 -1 0.16 10 -2 0.25 10 -3 0.33 10 -4 To correct for physical decay of technetium Tc 99m, the fractions that remain at selected intervals after the time of calibration are shown in Table 3 . Table 3 * Calibration Time Physical Decay Chart: Tc 99m, half-life 6.02 hours Hours Fraction Remaining Hours Fraction Remaining 0* 100.0 7 0.447 1 0.891 8 0.398 2 0.794 9 0.355 3 0.708 10 0.316 4 0.631 11 0.282 5 0.562 12 0.251 6 0.501

Spl Unclassified Section Table

Table 1
1 Kocher, David C: Radioactive Decay Data Tables, DOE/TlC-11026, 108, 1981

Principal Radiation Emission Data

Radiation

Mean % per Disintegration

Mean Energy (keV)

Gamma-2

89.07

140.5

Nursing Mothers

Nursing Mothers Technetium Tc 99m is excreted in human milk during lactation; therefore, formula-feedings should be substituted for breast-feedings.

Pediatric Use

Pediatric Use Safety and effectiveness in children have not been established.

Pregnancy

Pregnancy Animal reproduction studies have not been conducted with technetium Tc 99m medronate. It is also not known whether technetium Tc 99m medronate can cause fetal harm when administered to a pregnant woman or can affect reproductive capacity. Technetium Tc 99m medronate should be given to a pregnant woman only if clearly needed.

How Supplied

HOW SUPPLIED Kit for the Preparation of Technetium Tc 99m Medronate is supplied as kits of 10 reaction vials (NDC 65857-505-10). Each reaction vial contains a sterile, nonpyrogenic, nonradioactive lyophilized mixture of 20 mg medronic acid, 1 mg ascorbic acid, 0.13 mg (minimum) stannous fluoride, SnF 2 and 0.38 mg total tin, maximum (as stannous fluoride, SnF 2 ). The pH is adjusted with sodium hydroxide or hydrochloric acid prior to lyophilization. The vial does not contain a preservative. The contents of the vial are lyophilized and sealed under nitrogen at the time of manufacture. The pH of the reconstituted product is 5.4 to 6.8. Kit Contents 10 sterile reaction vials 20 pressure-sensitive labels for technetium Tc 99m medronate 1 package insert Storage Store the product as supplied at 20° to 25°C (68° to 77°F). After reconstitution store refrigerated at 2° to 8°C (36° to 46°F) ( see DOSAGE AND ADMINISTRATION ). Do not use and discard radiolabeled technetium Tc 99m medronate 6 hours after reconstitution.

Storage And Handling

Storage Store the product as supplied at 20° to 25°C (68° to 77°F). After reconstitution store refrigerated at 2° to 8°C (36° to 46°F) ( see DOSAGE AND ADMINISTRATION ). Do not use and discard radiolabeled technetium Tc 99m medronate 6 hours after reconstitution.

General Precautions

General The contents of the kit before preparation are not radioactive. However, after the sodium pertechnetate Tc 99m injection is added, adequate shielding of the final preparation must be maintained. Contents of the reaction vial are intended only for use in the preparation of technetium Tc 99m medronate and are NOT to be administered directly to the patient. Technetium Tc 99m medronate as well as other radioactive drugs, must be handled with care. Once sodium pertechnetate Tc 99m is added to the vial, appropriate safety measures should be used to minimize external radiation to clinical occupational personnel. Care should also be taken to minimize radiation exposure to patients in a manner consistent with proper patient management. To minimize the radiation dose to the bladder, the patient should be encouraged to drink fluids and to void immediately before the examination and as often thereafter as possible for the next four to six hours. Technetium Tc 99m medronate should be formulated within six (6) hours prior to clinical use. Optimal imaging results are obtained one to four hours after administration. Technetium Tc 99m medronate injection should be discarded six hours after reconstitution. The solution should not be used if cloudy. The components of the kit are sterile and pyrogen-free. It is essential to follow directions carefully and to adhere to strict aseptic procedures during preparation. The technetium Tc 99m labeling reactions involved in preparing the agent depend on maintaining the stannous ion in the reduced state. Any oxidant present in the sodium pertechnetate Tc 99m supply may, thus, adversely affect the quality of the radiopharmaceutical. Hence, sodium pertechnetate Tc 99m containing oxidants should not be employed. Radiopharmaceuticals should be used only by physicians who are qualified by training and experience in the safe use and handling of radionuclides.

Precautions

PRECAUTIONS General The contents of the kit before preparation are not radioactive. However, after the sodium pertechnetate Tc 99m injection is added, adequate shielding of the final preparation must be maintained. Contents of the reaction vial are intended only for use in the preparation of technetium Tc 99m medronate and are NOT to be administered directly to the patient. Technetium Tc 99m medronate as well as other radioactive drugs, must be handled with care. Once sodium pertechnetate Tc 99m is added to the vial, appropriate safety measures should be used to minimize external radiation to clinical occupational personnel. Care should also be taken to minimize radiation exposure to patients in a manner consistent with proper patient management. To minimize the radiation dose to the bladder, the patient should be encouraged to drink fluids and to void immediately before the examination and as often thereafter as possible for the next four to six hours. Technetium Tc 99m medronate should be formulated within six (6) hours prior to clinical use. Optimal imaging results are obtained one to four hours after administration. Technetium Tc 99m medronate injection should be discarded six hours after reconstitution. The solution should not be used if cloudy. The components of the kit are sterile and pyrogen-free. It is essential to follow directions carefully and to adhere to strict aseptic procedures during preparation. The technetium Tc 99m labeling reactions involved in preparing the agent depend on maintaining the stannous ion in the reduced state. Any oxidant present in the sodium pertechnetate Tc 99m supply may, thus, adversely affect the quality of the radiopharmaceutical. Hence, sodium pertechnetate Tc 99m containing oxidants should not be employed. Radiopharmaceuticals should be used only by physicians who are qualified by training and experience in the safe use and handling of radionuclides. Carcinogenesis, Mutagenesis, Impairment of Fertility No long-term animal studies have been performed to evaluate carcinogenic potential or whether technetium Tc 99m medronate affects fertility in males or females. Pregnancy Animal reproduction studies have not been conducted with technetium Tc 99m medronate. It is also not known whether technetium Tc 99m medronate can cause fetal harm when administered to a pregnant woman or can affect reproductive capacity. Technetium Tc 99m medronate should be given to a pregnant woman only if clearly needed. Nursing Mothers Technetium Tc 99m is excreted in human milk during lactation; therefore, formula-feedings should be substituted for breast-feedings. Pediatric Use Safety and effectiveness in children have not been established.

Learning Zones

The Learning Zones are an educational resource for healthcare professionals that provide medical information on the epidemiology, pathophysiology and burden of disease, as well as diagnostic techniques and treatment regimens.

Disclaimer

The drug Prescribing Information (PI), including indications, contra-indications, interactions, etc, has been developed using the U.S. Food & Drug Administration (FDA) as a source (www.fda.gov).

Medthority offers the whole library of PI documents from the FDA. Medthority will not be held liable for explicit or implicit errors, or missing data.

Drugs appearing in this section are approved by the FDA. For regions outside of the United States, this content is for informational purposes only and may not be aligned with local regulatory approvals or guidance.